Support 4D probe; recognize the probe through the ID code; switch the 4D push sign by way of the  electrical relay to make the 4D probe perform Typically. four.two.two Main board The most crucial board can be divided into two areas: front-close of the main board and again-conclude of the key board.

The consumer is not really permitted to open up the addresses and panel of your technique, neither Observe: machine disassemble is permitted. To ensure the process functionality and security, only Mindray engineers or engineers authorized by Mindray can complete upkeep.

three. Click the higher icon from the display (prompts “Update” if you set the cursor on it), the update stage is carried out, if fails, please deliver to R&D Heart. Determine six-one Principal Watch Routine maintenance Display 4. After update is profitable, restart the equipment right after ability off to create the info productive. HDD Partition ... Page ninety one structured report related details file and DICOM viewing computer software Report report template obd.bin OBD knowledge file Model.txt Model file Comment Remark file AnatomyImage Anatomical pictures Measurement Measurement library Doppler program and associated boot configuration most important file and plug-ins, distant desktop server Multilanguage string file exe videoplay.

This manual may possibly referring to information and facts safeguarded by copyright or patents and would not convey any license beneath the patent legal rights or copyright of Mindray, or of Many others. Web site nine: Statement

Slash off cable ties connecting the main device over the Display screen by diagonal cutting pliers, then pull out the corresponding cable plugs.

Mindray’s new era of black & white ultrasound technique, DP-50 Expert, is supplied with Innovative imaging technologies allowing deeper penetration and quicker graphic acquisition with bigger resolution.

Web site a hundred forty five Cleanse control panel Instruments: dry comfortable fabric, delicate soapy drinking water  Strategy:  Use dry tender fabric to clean the surface with the program. If the program is filthy, moisten the delicate fabric with gentle soapy h2o and wipe off any stains. Use dry tender cloth to eliminate any humidity and permit all difficult surfaces to absolutely air-dry. Webpage 146: Cleanse The Peripherals

Uncover better diagnostic info through prolonged watch of your anatomical construction on all convex probes.

Analyzing sufferers in today’s demanding natural environment necessitates innovative engineering that enables clinicians to create quick and self-confident diagnoses. The versatility from the DP-50 expands your scientific utility although assisting to strengthen client care.

five.2.3 Geometric positioning accuracy Longitudinal geometric positioning accuracy  Exam Step: Do adjustments as how in tests the maximum depth. Document the gap by 20mm Each individual segment within the longitudinal targets line using the measurement caliper; Pick out the worth with the greatest Dp50rb error (to 20mm), determine the precision using the components underneath The measurement caliper should be positioned within the higher fringe of the goal, Take note:... Web site 79 Transverse geometric positioning accuracy  Examination Phase: Deal with the scan floor in the phantom with water or few gel, gently contact the probe Using the scan surface area Modify the depth, building the transverse targets being shown while in the image. Modify the main target issue being posited beside the transverse targets (the normal just isn't distinct) Regulate parameters like acquire, TGC, building Every transverse targets to become clearly displayed. Page 80 5.five.2.four Blackout Area Check Step: Include the scan surface on the phantom with water or few gel, gently contact the probe Using the scan surface Regulate the depth at a decrease price, and established the main focus at the closest location to the scan area. Lower the worth of parameters like AP, Achieve right until the track record noise just might be found. Page eighty one: Software program Up grade And Servicing

Kebutuhan Anda akan alat profesional dan canggih yang dikombinasikan dengan pencitraan berkualitas dan kemudahan kegunaan kini dapat dipenuhi dengan sistem portabel baru Mindray DP-50Exp.

.. Page 161 The indicator on the printer Glitches happen when the adapter turns orange, and also the printing printer adapter runs. Disconnect and link the fails. ability adapter, ability over the printer adapter all over again, and confirm whether the printing is normal after 1 moment. Webpage 163: Electrical Safety Inspection

General performance Exam five.5.one Examination Process Resolution Detection depth Geometric Posture Precision Blackout spot exam Report the Test photos five.five.2 Test Written content The picture employed Here's just for reference, stay with the picture influence in the real Be aware: situation. Specifications: Display screen: set the contrast and brightness at the medical software value (or perhaps the default position) Operation ecosystem: dark room, simulating the scientific application surroundings. Page 76 In issue which the transverse resolution tests targets are horizontally shown, record the nominal distance of two targets that can be Evidently regarded. Repeat the Procedure over to the transverse resolution testing targets at other depths. As proven in figure beneath. Axial resolution ... Website page seventy seven When utilizing the convex probe, maintain the transverse resolution testing targets to notice: be shown close to the midline. When employing a linear probe with steer perform, usually do not turn on the steer function when conduct the transverse resolution take a look at. Zoom in the location wherever the targets Positioned if important. The diameter on the goal point at a specific depth is equal to the transverse resolution within the depth. Site 78 5.

Mindray respects your privacy. We use cookies for several different functions, like Web site operation, increasing your expertise of our Site, setting up integration with social websites and aiding target our promotion.

Notice: Make sure that you could possibly pull out and drive into your IO board sockets vertically when pulling out and assembling the IO board assembly about the place with the deal with. Take out panhead screws with washers M3X6 (4) used to safe the IO board, after which you can you could consider out the IO board. Webpage 116 Take away panhead screws with washers (2 M3X6) which might be preset on the key device, choose out the higher baffle and afterwards pull out two cables connecting the primary board to the power module assembly.